• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP Back to Search Results
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device has been returned to animas. Evaluation has not yet been completed. When evaluation is complete a supplemental report will be filed. No conclusion can be made at this time.
 
Event Description
On (b)(6) 2017, the reporter contacted animas, alleging a display (dim/fading/color spectrum) issue. There was no indication that the device caused or contributed to an adverse event. This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
 
Manufacturer Narrative
The device has been returned and evaluated by product analysis on 05/02/2017 with the following findings: during investigation, the pump was powered on to a dim and pink display. Unrelated to the original complaint, the battery compartment was cracked at the top left, and lower left corners of the grip pad.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANIMAS VIBE
Type of DeviceINSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6510113
MDR Text Key73509335
Report Number2531779-2017-08209
Device Sequence Number0
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 03/29/2017
2 Devices were Involved in the Event: 1   2  
0 Patients were Involved in the Event:
Date FDA Received04/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received03/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-