The scope was returned to olympus for evaluation; however, the device evaluation has not yet begun.
The scope will be sent to an independent laboratory for culture testing and ethylene oxide (eto) sterilization.
As part of our investigation, an olympus endoscopy support specialist (ess) was dispatched to the user facility on april 17, 2017 to observe the facilities reprocessing practice and to provide a reprocessing training.
The ess found reprocessing deviations with leak testing, manual cleaning, and high level disinfection of the scope and the scope connectors.
The ess educated the user facility on the following: leak testing should be completed after each use.
Proper dilution of flexclean according to the mixing instruction must be followed.
Proper brushing of scope has to be performed and worn brushes have to be replaced.
Proper disinfection and cleaning of reusable brushes has to be performed.
Properly testing the aldahol solution with test strips and logging the test results.
In addition to educating the user facility staff of proper reprocessing practices, the ess provided the staff a cleaning guide wall poster information.
Based on several reprocessing deviations that the ess found during the on-site visit, the cause of the reported patient infections is likely attributed to insufficient reprocessing of the scope.
The reprocessing manual for use provides several warning and caution statements in an effort to prevent patient infection and cross contamination.
¿failure to properly clean and high-level disinfect or sterilize endoscopic equipment after each procedure may compromise patient safety.
To minimize the risk of transmitting diseases from one patient to another, after each procedure the endoscope and its ancillary equipment must undergo thorough manual cleaning followed by high-level disinfection or sterilization.
Reprocess not only the external surface of the endoscope but also all channels.
All channels of the endoscope must be cleaned and high-level disinfected or sterilized during every reprocessing cycle, even if the channels were not used during the previous patient procedure.
If the endoscope is not cleaned meticulously, effective disinfection or sterilization may not be possible.
Clean the endoscope and accessories thoroughly before disinfection or sterilization to remove microorganisms and organic material that could reduce the efficacy of disinfection or sterilization.
Otherwise, insufficient cleaning and disinfection or sterilization of the endoscope may pose an infection control risk to the patient and/or operators performing the next procedure with the endoscope.
¿.
|
The scope was sent to an independent laboratory for culture testing.
The scope tested negative for growth.
The scope was then ethylene oxide (eto) sterilized and returned to olympus for a physical device evaluation.
A visual inspection was performed on the scope and noted brownish stains on the distal end cover and objective lens glue.
In addition, a boroscope was used inspect the biopsy channel and found tear marks near the distal end and bending section side of the scope.
There were no signs for foreign materials found on the bending section cover, bending section cover glue and insertion tube.
The scope passed leak testing.
The scope was serviced and returned to the user facility.
Based on the tear marks found on the biopsy channel and brownish stain found on the distal end cover, user handling and improper maintenance could not be ruled out as contributory factors to the reported event.
|