• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CLEO® 90 INFUSION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CLEO® 90 INFUSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 21-7220-24
Device Problems Device Alarm System (1012); Bent (1059); Partial Blockage (1065)
Patient Problem Hyperglycemia (1905)
Event Date 03/23/2017
Event Type  Injury  
Manufacturer Narrative
Customer has not returned the device to the manufacturer for device evaluation.If the device becomes available and is returned and evaluated, the manufacturer will file a follow-up report detailing the results of the evaluation.
 
Event Description
It was reported that a cleo® 90 infusion set was in use with a patient when the patient observed multiple occlusion alarms (alarm number 2 and 26 in pump history) during a bolus.Troubleshooting determined that there was a blockage at the site due to the cannula being bent and creased at the base.The infusion set had been in use for 48 hours.The patient used the insulin type humalog.Due to the event, the patient's blood glucose levels were reported to be above 400mg/dl.The patient administered a bolus via the infusion pump to address the high blood glucose.The patient planned to change supplies and resume insulin therapy.No permanent injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEO® 90 INFUSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS HEALTHCARE MFG
s.a. de c.v. ave calidad no. 4
parque industrial internaciona
tijuana, 22425
MX   22425
Manufacturer Contact
lisa perz
6000 nathan lane north
minneapolis, MN 55442
7633833074
MDR Report Key6522726
MDR Text Key73730985
Report Number3012307300-2017-00891
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586028335
UDI-Public10610586028335
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042172
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/19/2021
Device Catalogue Number21-7220-24
Device Lot Number76X172
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age10 YR
-
-