Model Number BE-HQV 33800
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/07/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Maquet cardiopulmonary (b)(4) is aware of similar complaints for this product.Similar products, showing a similar malfunction, have been tested.The tested products had the same leakage at the same position.This could be determined as bonding failure.Visual inspection, tightness test acc.To lv 203 and tightness test with roller pump have been performed.Results: during tightness test acc.To lv 203 the exact position of the leak in the arterial filter between the cover and the housing could be detected (improper ultrasonic welding).During testing with roller pump leakage at quart could also be detected but the exact position was not found during this kind of test.The existing liquid collected at the lowest point of the filter and drained.Most possible root cause could be the bad bonding between cover and filter body.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
During priming, a leak occurred at the side of the housing of the arterial filter.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BE-HQV 33800#COMPLETE PACK
Type of Device
FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6523184
MDR Text Key74012542
Report Number8010762-2017-00131
Device Sequence Number1
Product Code DTM
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K090518
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/01/2018
Device Model NumberBE-HQV 33800
Device Catalogue Number70101.8883
Device Lot Number92203567
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2017
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-