Catalog Number 42544
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2017
Event Type  malfunction  
Manufacturer Narrative
The product testing is in progress.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported to medtronic neurosurgery that the reservoir's flexibility was found to be poor at the pre-operation test.According to the report, the reservoir could not be easily filled.
 
Manufacturer Narrative
The returned valve was patent.It met the requirements for reflux, pressure-flow, preimplantation and leak testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ a review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL VALVE, BURR HOLE, MEDIUM
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis,mn, CA 55432
7635260594
MDR Report Key6529360
MDR Text Key74019519
Report Number2021898-2017-00235
Device Sequence Number1
Product Code JXG
UDI-Device Identifier00643169466913
UDI-Public00643169466913
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K831678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 03/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number42544
Device Lot NumberD84234
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-