• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES MONUMENT 11MM/130 DEG TI CANN TFNA 170MM - STERILE; ROD, FIXATION, INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES MONUMENT 11MM/130 DEG TI CANN TFNA 170MM - STERILE; ROD, FIXATION, INTRAMEDULLARY Back to Search Results
Catalog Number 04.037.142S
Device Problems Fitting Problem (2183); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/03/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information is not available for reporting.(b)(4).The subject device is not expected to be returned to the synthes manufacturer for evaluation.(b)(6).The investigation could not be completed; no conclusion could be drawn, as no product was received.A device history record review was performed for the reported subject device lot.Manufacturing location: synthes (b)(4).Date of manufacture: dec 21, 2015.Expiration date: dec 31, 2025.The review showed that there were no issues during the manufacture of the product lot that would contribute to this complaint condition.No non-conformances were generated during the production of the subject device.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Synthes (b)(4) reported an event in (b)(6) as follows: it was reported that during surgery to implant a trochanteric fixation nail anitrotation (tfna) nail, tfna blade, unknown end cap, and unknown screw on (b)(6) 2017 to treat a trochanteric femoral fracture, even though a ¿click¿ sound could be heard, the locking mechanism could not be locked.The surgeon was able to insert the end cap and the hardware was left in the patient.There was a surgical delay of unknown duration due to the reported event.The patient¿s postoperative outcome is unknown.Concomitant part: 1 x end cap, part: unknown this report is 1 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11MM/130 DEG TI CANN TFNA 170MM - STERILE
Type of Device
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer (Section G)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6530524
MDR Text Key74152952
Report Number1719045-2017-10355
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131548
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.037.142S
Device Lot Number9961793
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/03/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/21/2015
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNKNOWN END CAP, QTY. 1
-
-