MAUDE Adverse Event Report: AESCULAP AG FILTER RETENTION PLATE; ACCESSORIES CONTAINER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG FILTER RETENTION PLATE; ACCESSORIES CONTAINER SYSTEM Back to Search Results
Model Number JK100
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Manufacturing site evaluation: evaluation on-going.
 
Event Description
(b)(6).It was reported that the filter holder is detached.All med watch submissions related to this report are: 9610612-2017-00207, 9610612-2017-00208, 9610612-2017-00209, 9610612-2017-00210, 9610612-2017-00211.
 
Manufacturer Narrative
Investigation: the investigation was carried out visually.All seals are damaged or fallen off.All plates have the same maintenance date, october 2014 at ats in (b)(4).The silicone is brittle, as it is already several years old.Batch history review: the product does not require batch management; a review of the device quality and manufacturing history records is not possible.Conclusion and root cause: based on the information available as well as a result of our investigation the root cause of the failure is most probably related to an insufficient maintenance of the device.Rational: the maintenance of the retention plates has been carried out insufficiently.The silicon inlays are brittle due to normal aging.There is no maintenance instruction for retention plates available.Instead of using glue, the plates should have been replaced.Corrective action according to sop (b)(4) a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILTER RETENTION PLATE
Type of Device
ACCESSORIES CONTAINER SYSTEM
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key6531449
MDR Text Key74169045
Report Number9610612-2017-00207
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K792558
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberJK100
Device Catalogue NumberJK100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2017
Is the Reporter a Health Professional? No
Distributor Facility Aware Date04/17/2017
Date Manufacturer Received02/21/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-