Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Death (1802); Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226)
Event Date 03/22/2017
Event Type  Death  
Event Description
It was reported that during a cryoablation procedure the patient died.The patient experienced massive spontaneous blood pressure decrease due to tamponade, which led to asystole.Pericardial drainage was performed three times and attempts were made to revive the patient, unsuccessfully.Also, an echocardiography (echo) was performed.The patient expired and no autopsy was performed.
 
Manufacturer Narrative
The data files for the date of the event were returned and analyzed.The files show five injections were performed with balloon catheter 2af283, lot 76108, without issue.The console was properly controlling flow and pressure.No product malfunction was alleged; a known clinical issue was encountered during the procedure which led to death of the patient.The physical products involved during the procedure were not returned for analysis.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6534618
MDR Text Key74094154
Report Number3002648230-2017-00199
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 04/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/30/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death; Life Threatening; Required Intervention;
Patient Age73 YR
-
-