MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-27
Device Problems Detachment Of Device Component (1104); Use of Device Problem (1670)
Patient Problem Foreign Body In Patient (2687)
Event Date 04/04/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4). The g5 system is associated with product code pqf.
 
Event Description
Dexcom was made aware on (b)(6) 2017, that on (b)(6) 2017, of a detached sensor wire that was retained in the patient's skin. The sensor was inserted at the abdomen on (b)(6) 2017. Patient's mother reported that this was her first insertion. The sensor felt normal after insertion. Then the sensor failed to pair or calibrate. Patient's mother could not see any portion of the sensor wire above the patient's skin. Patient's mother palpated the insertion area and she felt something was there. She also reported visually seeing a blood clot at the insertion area. Patient has not been seen by a clinician. There was no reported sign of infection. At the time of contact, patient is okay. No further event or patient information is available. No product or data was provided for investigation. The reported event could not be confirmed. The root cause could not be determined. The transmitter was removed prior to removing the sensor pod. Labeling indicates: do not remove the transmitter before removing the sensor pod from your body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6538148
MDR Text Key74210468
Report Number3004753838-2017-35670
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000019
UDI-Public(01)00386270000019(241)9500-27(10)NI(17)NI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation
Type of Report Initial
Report Date 04/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9500-27
Device Catalogue NumberSTS-GL-011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 05/02/2017 Patient Sequence Number: 1
-
-