• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-01
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Sepsis (2067)
Event Date 07/13/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
The recipient reportedly experienced a wound gap with sepsis and device migration.On (b)(6) 2016, the recipient underwent resuturing, however, the implant remained visible through the incision.On (b)(6) 2017, the recipient underwent a flap rotation revision surgery.The recipient was treated with iv antibiotics for 7 days, however, the issue did not resolve.The recipient's device was explanted.The recipient was not reimplanted.The recipient's implant site is healed.
 
Manufacturer Narrative
(b)(4).The device passed both the external visual and photographic imaging inspections.System lock was verified.The device passed all of the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
rachael perez
28515 westinghouse place
valencia, CA 91355
6613627734
MDR Report Key6539282
MDR Text Key74252950
Report Number3006556115-2017-00223
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/31/2017
Device Model NumberCI-1500-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2017
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-