CAREFUSION/BD RESUS, ADLT W/MASK, 40" TBG, 6/CS CARDIOPULMONARY RESUSCITATION AID KIT Back to Search Results
Catalog Number 2K8005
Device Problem Failure to Disconnect (2541)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/09/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
Customer reported the following via medwatch (b)(4). Customer reported that the "patient was being bag/mask ventilated; once endotracheal tube placed, attempted to remove mask from the resuscitation bag and the mask was unable to be removed from the resuscitation bag".
 
Manufacturer Narrative
Unfortunately the sample was never returned to vyaire for further evaluation. At this time the reporter unable to disconnect failure mode cannot be confirmed without the device sample. However based on similar complaints a probable root cause is considered to be the mirror finish surface on the elbow making the mask very difficult to remove. A new textured surface finish has been implemented on the elbows allowing the mask to be easily removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameRESUS, ADLT W/MASK, 40" TBG, 6/CS
Type of DeviceCARDIOPULMONARY RESUSCITATION AID KIT
Manufacturer (Section D)
CAREFUSION/BD
cerrada vía de la producción
no. 85 parque industrial
mexicali baja california norte
MX
Manufacturer (Section G)
CAREFUSION/BD
cerrada vía de la producción
no.85 parque industrial
mexicali baja california norte
MX
Manufacturer Contact
mindy faber
75 north fairway drive
vernon hills, IL 60061
MDR Report Key6541335
MDR Text Key74318572
Report Number8030673-2017-00332
Device Sequence Number1
Product Code OEV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
ENFORCEMENT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 07/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number2K8005
Device Lot Number0001030580
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/14/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 05/03/2017 Patient Sequence Number: 1
-
-