Model Number N/A
Device Problem Unintended Collision (1429)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2017
Event Type  malfunction  
Manufacturer Narrative
The device br16018 has not been evaluated yet for investigation purpose.Once the evaluation will be performed, a follow-up medwatch report will be submitted.
 
Event Description
During surgery the device robot arm collided with the mayfield head frame.The collision was detected by the software and device did shutdown.Surgery was pursued after a device restart.
 
Manufacturer Narrative
Investigation into this event indicated that the reason for the collision is user related.User should have stopped the robot arm movement by releasing the vigilance device to avoid the collision.The ifu contains necessary instructions about the vigilance device usage, and the surgeon who was operating the device when the event occurred has been trained to the use of the rosa brain device.The device shutdown that followed the collision is a normal system behaviour: the shutdown is a safety measure, which immediately stops the device and limits the impact of the collision.Corrected data: date of this report.Date received by manufacturer.If follow-up, what type.Evaluation code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA BRAIN
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key6544149
MDR Text Key74426002
Report Number3009185973-2017-00604
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK151359
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 11/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-