Catalog Number 72200616S
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/18/2017
Event Type  malfunction  
Manufacturer Narrative
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed. A relationship, if any, between the subject device and the reported event could not be determined. A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Manufacturer Narrative
The device was received for evaluation. A visual inspection was performed on the product and no issue was observed. There was a relationship found between the returned device and the reported incident. Product failed functional testing with motor stall error and overheating. Cause of overheating and errors is a corroded motor/gearbox. The motor/gearbox assembly could not be removed from the housing for further assessment due to corrosion. The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly. A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the mdu was overheating. This was noticed before the procedure. No patient or staff impact was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVC REPL,MDU, HAND CNTRL, PWRMX
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key6566176
MDR Text Key138085572
Report Number1643264-2017-00268
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 01/15/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-