In trouble-shooting, the medtronic representative reviewed the exam archive and found that the surgeon was using a cta with a large field of view (fov) that included the sternum.
The tracer appeared shifted laterally, since there were no points shown collected on the nose, only points showing next to the nose along the cheek.
It is suspected the initial blue dots appeared shifted as well and were not corrected and accuracy was not confirmed following tracer.
- return requested.
No parts have been received by manufacturer for analysis.
- on 04/19/2017 a medtronic representative performed a navigation system check-out.
Hardware and instruments testing passed.
- on 05/02/2017 software investigation completed.
Findings are that analysis of patient logs shows the patient exam was not to protocol and tracer pattern was not adequate.
Software is functioning as designed.
- medtronic representative reported the site has received additional training in proper protocol for patient exam and tracing patterns.
|
The physician in the case reported that while in a cranial biopsy procedure with the navigation system was inaccurate.
A sample was taken, and sent for testing.
However, when the biopsy was reviewed, it was healthy brain tissue.
There was no delay to the case and the patient is scheduled for revision surgery with another navigation system on (b)(6) 2017.
A medtronic representative reported that while in a cranial biopsy procedure the surgeon alleged an inaccuracy occurred.
No specific measurement, or direction, of the alleged inaccuracy was provided.
A sample was taken, and sent for testing, however, when the biopsy was reviewed, it was found to be healthy brain tissue.
There was no delay to the case and the patient is scheduled for revision surgery with another navigation system on (b)(6) 2017.
|