SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problems Overheating of Device (1437); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/21/2017
Event Type  malfunction  
Manufacturer Narrative
Device investigation narrative - complaint of overheating and motor stall error was confirmed. Cause of overheating and errors is a corroded motor/gearbox. The motor/gearbox was removed from the housing. The gearbox was removed from motor. The cable assembly and motor tested good. The gearbox was found to be jammed and corroded internally. A review of the manufacturing records shows that this unit was released to distribution on or about (b)(6) 2015. The root cause of this event was identified to be associated with corrosion of the motor and gearbox assembly. A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing. Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved. (b)(4).
 
Event Description
It was reported the motor drive unit hand cntrl pwrmx e generated an error code and the handle got extremely warm. A backup device was readily available. There were no delays or patient injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key6572317
MDR Text Key75431740
Report Number1643264-2017-00272
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation
Type of Report Initial
Report Date 05/17/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-