Model Number 24502082
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/19/2017
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment and on coils.Internally, no signs of foreign substance, burning, melting or charring were observed.Additional testing supports that the misplacement of a battery can cause battery leakage.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2017 to report leaking fluid within the battery compartment after using 6 aa batteries to power on her purely yours breast pump 2 days ago.Customer reports no contact with this fluid and denies any injury or burn.She states hearing an unusual sound from the pump base, stopped pumping and found white fluid leaking from some batteries that appeared expanded.She also states the pump base became warmer than usual while operating it with the ac adapter.Customer was overnight shipped a replacement pump base.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key6572882
MDR Text Key75419853
Report Number3009974348-2017-00266
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 04/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24502082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received04/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age31 YR
-
-