ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100H-12
Device Problem Device Operates Differently Than Expected (2913)
Patient Problems Pain (1994); Increased Sensitivity (2065)
Event Date 04/27/2017
Event Type  malfunction  
Event Description
The recipient reportedly experienced a hit to the head followed by loss of sound and loss of lock.External equipment was exchanged, however, the issue is not resolved.The recipient was recommended non-use of the device.The recipient is having pain and sensitivity at the implant site without device use.Revision surgery will be scheduled.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
The external visual inspection revealed that the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device failed the mechanical test performed.The failure of this device is attributed to excessive moisture through a gross leak at the case-band braze, which is consistent with the trauma reported.This ultimately caused this device to cease functioning.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
pamela campo
28515 westinghouse place
valencia, CA 91355
6613627624
MDR Report Key6577359
MDR Text Key75663335
Report Number3006556115-2017-00262
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Audiologist
Type of Report Initial,Followup,Followup,Followup
Report Date 04/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberAB-5100H-12
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2017
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
-
-