Catalog Number 72201500
Device Problems Device Operates Differently Than Expected (2913); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/25/2017
Event Type  malfunction  
Event Description
It was reported that during the ankle procedure, the motor was running but the blades were not turning and the device got hot.No reported patient injuries.No other complications were noted.
 
Manufacturer Narrative
Complaint of overheating was confirmed.Mdu failed for hand piece blade stall error as well as overheating.Cause of errors and overheating is a corroded motor/gearbox pn: (b)(4).The motor/gearbox was removed from the housing.The gearbox was removed from motor and motor tested good.The root cause of this event was identified to be associated with corrosion of the gearbox assembly.A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history record was performed which confirmed no inconsistencies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS POWERMINI, WITH HAND CONTROLS
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
MDR Report Key6581109
MDR Text Key75796007
Report Number1643264-2017-00281
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72201500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-