Model Number 200XDR
Device Problem Overheating of Device (1437)
Patient Problem Burn, Thermal (2530)
Event Date 04/25/2017
Event Type  malfunction  
Manufacturer Narrative
During a test run of the product the heat up was reproducible and it showed that the handpiece was running out of specification as it needed a too high current. This is a sign that the bearings are not running smooth anymore due to wear process. In addition it was found that the push button has groove marks on the inner side. This shows that it has been pressed while the handpiece has been spinning. This causes unusual friction and hence a quick heat up of the head. To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: caution: burning hazard from hot instrument head or hot instruments cover. If the instrument overheats, burns may arise in the oral area. ->never contact soft tissue with the instrument head or instrument cover the following guidelines must be observed to ensure save use of the electrically driven contra-angle handpieces: >the service instructions for contra-angle handpieces must be precisely following when using kavo spray or quattrocare care systems. > before each use, the contra-angle handpiece must be checked for external damage. > before each use, perform a test run with the contra-angle handpiece, and watch for atypical heating and unusual noise and vibration. > immediately stop using contra-angle handpieces that act unusual. > never press the pushbutton during operation. This also includes lifting the cheek or tongue! to ensure proper function, the medical device must be set up according to the reprocessing methods described in the kavo instructions for use, and the care products and care systems described therein must be used. Kavo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device. This service interval depends on the frequency of use and should be adjusted accordingly.
 
Event Description
Dr. (b)(6) on (b)(6) 2017 stated while working tooth# 1, 2, and 3 on (b)(6) 2017 during teeth extraction procedure, patient was burned on right cheek and right side of tongue. Burn diameter to cheek was quarter in size and size of dime on tongue. Patient was given an over the counter ointment. No medical care necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOMFORTDRIVE 200XDR
Type of DeviceDENTAL ELECTRICAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, bw 88400
GM 88400
Manufacturer (Section G)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, bw 88400
GM 88400
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach / riss, bw 88400
GM   88400
735156
MDR Report Key6607504
MDR Text Key76504849
Report Number3003637274-2017-00037
Device Sequence Number1
Product Code EKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080677
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial
Report Date 06/02/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/02/2017
Is this a Product Problem Report? Yes
Device Operator Dentist
Device Model Number200XDR
Device Catalogue Number1.007.3570
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-