• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/09/2017
Event Type  malfunction  
Manufacturer Narrative
The hospital biomed performed a checkout of the equipment and a review of the logs.The evap was replaced.Patient information could not be obtained after multiple attempts.Attempts were made as follows: 5/26/17 phone call, 5/31/17 phone call, 6/2/17 phone call.(b)(4).
 
Event Description
The hospital reported the clinician turned the vaporizer to 1%, and the gas module reportedly indicated 5%.It was further noted the patient had a drop in blood pressure.The clinician switched to intravenous anesthetic to complete the case.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI
Manufacturer Contact
john szalinski
540 w. northwest highway
barrington, IL 
MDR Report Key6621698
MDR Text Key76972402
Report Number2112667-2017-01053
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberSEE BLOCK H10
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-