• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.BRAUN SURGICAL SA DAFILON BLUE 1 (4)150CM HRN50 LOOP(M); SUTURES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA DAFILON BLUE 1 (4)150CM HRN50 LOOP(M); SUTURES Back to Search Results
Model Number B0934028
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Manufacturing site evaluation: evaluation on-going.
 
Event Description
Country of complaint: (b)(6).It was reported that the needle detached from the thread.
 
Manufacturer Narrative
Samples received: 1 open pouch.Analysis and results: there are no previous complaints of this batch.We manufactured and distributed in the market (b)(4) units of this batch.There are no units in our stock.We have received one empty sample (only first pack is available).However, without any closed sample we cannot carry out an analysis in order to take a decision.Final conclusion: in spite of receiving an empty sample, without closed samples a suitable analysis cannot be performed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 1 (4)150CM HRN50 LOOP(M)
Type of Device
SUTURES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
Manufacturer (Section G)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP   08191
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key6634896
MDR Text Key77457392
Report Number3003639970-2017-00295
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K151165
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/30/2019
Device Model NumberB0934028
Device Catalogue NumberB0934028
Device Lot Number614331
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2017
Is the Reporter a Health Professional? No
Distributor Facility Aware Date05/26/2017
Date Manufacturer Received05/18/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-