The customer stated that they received erroneous results for three patient samples tested for bilt3 bilirubin total gen.3 (tbil) on a cobas 8000 c 702 module (c702).Of the three samples, one had an erroneous result that was reported outside of the laboratory.The sample initially resulted as 0 umol/l and this value was automatically released to the lis and the requesting doctor.The doctor realized that the result was erroneous and requested repeat testing.The repeat result was 236.7 umol/l and this value was reported outside of the laboratory.No adverse events were alleged to have occurred with the patient.The tbil reagent lot number was 201416.The reagent expiration date was asked for, but not provided.Tbil calibrations are performed daily and have been very steady.Erroneous low results for the two other patient samples were found by the customer on (b)(6) 2017.While investigating this occasion, the field service engineer found that the customer was running the samples in the same rack in micro-cups on top of false bottom tubes.The engineer also found that the sample probe was dirty, possibly with gel.He replaced the probe and re-aligned it.
|