One catheter was returned.An injector cap was attached to the luer adapter, which was not fully seated in the overmolded extension of the catheter.There was a gap of a few millimeters.The luer adapter was removed.The posts had broken off of the luer adapter.A definitive root cause for this complaint could not be determined.However, based on the fact that the posts had broken off the luer adapter, it is possible that the detachment was caused by the application of undue force to the luer adapter during the attempted removal of the injector cap.First picc catheters are 100% in-process inspected at various times during manufacture.Catheters are also leak and flow tested to ensure their integrity.Additionally, the instructions for use indicate several ways users can mitigate the occurrence of catheter damage.
|