• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER Back to Search Results
Model Number ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Device Problem Device Handling Problem (3265)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/09/2017
Event Type  malfunction  
Manufacturer Narrative
Siemens healthcare diagnostics inc.(siemens) has determined that the cause of the needlestick injury to the operator's finger on the advia 2120 hematology system with dual aspirate autosampler is the operator's technique when performing the centering collar maintenance on the instrument.The advia 120/ 2120/ 2120i hematology system operator's guide states the following: "to avoid personal injury and exposure to a potential biohazard, you must cover the needle with the red needle cover immediately after you remove the entering collar." the operator failed to follow the operator's guide instructions and did not cover the autosampler needle with the red needle stick cover immediately after removing the centering collar.The instrument is performing according to specifications.No further evaluation of this device is required.
 
Event Description
An operator sustained a needlestick injury to the finger while performing the centering collar maintenance procedure on the advia 2120 hematology system with dual aspirate autosampler.The operator went to health service in the same hospital where she sustained her injury.The operator refused the retroviral treatment offered by the hospital.There are no known reports of adverse health consequences due to the operator sustaining a needlestick injury while performing the centering collar maintenance procedure on the advia 2120 hematology system with dual aspirate autosampler.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Type of Device
ADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD.
chapel lane, swords, co.
registration number: 8020888
dublin,
EI  
Manufacturer Contact
elizabeth bernasconi
511 benedict ave
tarrytown, NY 10591
9145242495
MDR Report Key6662785
MDR Text Key78238188
Report Number2432235-2017-00372
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K102644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA 2120 HEMATOLOGY SYSTEM WITH DUAL ASPIRATE AUTOSAMPLER
Device Catalogue Number10313419
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/22/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-