Catalog Number 82-3100
Device Problem Device Operational Issue (2914)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Udi: product made prior to compliance date, gtin unavailable complaint sample was not returned to codman and no lot number information has been provided; therefore, an evaluation of the device could not be performed and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.Complaint will be closed at this time as 'no complaint sample returned to codman for evaluation'.If the complaint sample becomes available, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar complaints.
 
Event Description
The 1049 describe event valve needs to be explanted due to dysfunction.Valve was no longer adjustable.
 
Manufacturer Narrative
Updated udi: product made prior to compliance date, gtin unavailable; (b)(4).Upon completion of the investigation it was noted that the images were taken of the valve ¿as received¿.The valve was visually inspected; no defects were noted.The position of the cam when valve was received was 70 mmh2o.The valve was hydrated.The valve was tested for programming with programmer 82-3126 with serial (b)(4), the valve passed the test.The valve was flushed, the valve passed the test no occlusion was noted.The valve was leak tested, no leaks noted.The catheter was irrigated, no occlusions were noted.The valve was reflux tested, the valve passed the test.The valve was dried.The valve was then pressure tested, the valve passed the test.Review of the history device records confirmed the valve product code 82-3100 with lot cncc3p, conformed to the specifications when released to stock on the 28th march 2012.No root cause could be determined, as the problem reported by the customer could not be duplicated.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM & COMPS
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
325 paramount drive
raynham MA 02767
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key6667050
MDR Text Key78391687
Report Number1226348-2017-10442
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K974739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Catalogue Number82-3100
Device Lot NumberCNCC3P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received07/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-