• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PULSE GEN MODEL 103; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PULSE GEN MODEL 103; GENERATOR Back to Search Results
Model Number 103
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2017
Event Type  malfunction  
Event Description
It was reported that high impedance was found on the patient's device.The patient's vns system had been implanted for about five months.The device history records of the generator and lead were reviewed, and the devices conformed to all specifications prior to release.No further relevant information has been received to date.
 
Event Description
The patient had exploratory surgery, and the lead pin was identified to not be inserted fully.Once the lead pin was re-inserted, the high impedance resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 103
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6668807
MDR Text Key78443955
Report Number1644487-2017-04046
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodePE
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/21/2018
Device Model Number103
Device Lot Number5087
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received06/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age3 YR
-
-