Model Number 2420-0007
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/25/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Although requested, the affected product has not been received. A follow up report will be submitted with investigation results should the devices be received for evaluation.
 
Event Description
Received a copy of the customer's medwatch report from fda, which states ¿ iv alarm was reading occluded, the side channel was opened, and found the iv tubing had expanded into a large shaped circle inside the chamber, the iv rate was at 10cc/hour, the tubing had been in place approximately 9 -10 hours. " the customer reported that the clinician opened the pump module door due to an occlusion alarm and found the iv tubing had ballooned in the pump segment. There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameALARIS® PUMP MODULE ADMINISTRATION SET
Type of DeviceSET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
stephen bilello
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key6669675
MDR Text Key78470629
Report Number9616066-2017-00881
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number2420-0007
Device Catalogue Number2420-0007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/31/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/27/2017 Patient Sequence Number: 1
Treatment
8015,8100
-
-