• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
The customer reported a high blower temperature (e/c-1122) occurred.The unit was being used on a patient at the time the reported issue occurred; however, there was no patient harm.
 
Manufacturer Narrative
During further investigation (fi), a visual inspection of the blower assembly outer surface revealed no evidence of damage or contamination.The blower assembly end cap was removed and a visual inspection of the impellers and surroundings did not reveal any signs of rubbing, damage, or contamination.Before applying power to the blower assembly the lm20 temperature sensor output was verified.The blower assembly was installed into an fi test ventilator.The ventilator was booted up in normal ventilation mode and the ventilator remained operational with no errors detected.The test ventilator booted up and delivered breaths on ac.The device was allowed to run a minimum of 4 hours on normal ventilation mode and no error codes were generated.The fi test ventilator with the returned blower assembly installed was allowed to run another minimum of 4 hours on diagnostic mode with the restrictor removed from the gas outlet port.The test ventilator booted up and delivered breaths on ac and did not generate any errors while operating in diagnostic mode.Blower temperature readings remained below the nominal = 55¿.The air flow accuracy test was performed and passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
robert corning
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key6670023
MDR Text Key78668345
Report Number2031642-2017-02075
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 06/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-