MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS 6000 C (501) MODULE; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS 6000 C (501) MODULE; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number C501
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/15/2017
Event Type  malfunction  
Manufacturer Narrative
Unique identifier (udi)#: (b)(4).This event occurred in (b)(6).
 
Event Description
The customer obtained a questionable low test result for one patient sample using the crpl3 c-reactive protein gen.3 (crpl3) assay on the cobas 6000 c (501) module.All results are in units of mg/l, and all were released outside of the laboratory.The initial crpl3 result was 1.This result was automatically sent to the physicians.The customer noticed that the results for other tests for this patient had unspecified alarms, so the customer repeated the sample.The crpl3 result was 334.7.No other result data was provided.There was no allegation that an adverse event occurred.The crpl3 reagent lot number and expiration date were not provided.Investigation activities are ongoing.
 
Manufacturer Narrative
Quality control and precision testing information were acceptable.A specific root cause could not be identified.Additional information for further investigation was requested but was not provided.The most likely root cause for the non-reproducible low discrepant result was a partially blocked sample probe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 6000 C (501) MODULE
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6677236
MDR Text Key78729821
Report Number1823260-2017-01370
Device Sequence Number0
Product Code DCN
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K060373
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC501
Device Catalogue Number04745914001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age12 MO
-
-