It was starting in the middle of the night, the patient experienced ongoing occlusion alarms (alarm number in pump history: 2) during use of a cleo® 90 infusion set.
Through troubleshooting with the distributor, it was determined the blockage was likely located at the site.
The patient's blood glucose levels were 245 mg/dl at the time of the event.
To address the high blood glucose levels, the patient administered a bolus and manual injection.
The patient did not test for ketones.
No permanent injury was reported.
See mfr: 3012307300-2017-01402.
|