EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION Back to Search Results
Model Number PE075F5
Device Problem Pacing Problem (1439)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/08/2017
Event Type  malfunction  
Manufacturer Narrative
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe at gate valve was returned for evaluation.Continuity testing found that a short condition occurred around the catheter tip.A cut down of the catheter revealed that lead wires were not insulated inside the catheter tube at 0.8 inch proximal from the proximal electrode.This condition allows contact to be made with both wires and cause a short condition.The balloon inflated clear and concentric and remained inflated for 5 minutes without leakage.No visible damage or abnormality to the catheter body, balloon, windings or returned syringe was observed.Balloon inflation testing was performed using returned syringe with 1.3 cc air by holding the balloon under water.Visual examinations were performed under microscope at 20x magnification and with the unaided eyes.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of pacing issue was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Swan-ganz pacing thermodilution (td) catheters serve as diagnostic and therapeutic tools in the management of critically ill patients.There are multiple failure modes that may require the exchange of a pacing catheter.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.Stretching, kinking, or forceful wiping of the catheter may result in damage.After stable pacing has been confirmed, the proximal end of the catheter should be secured to the insertion site to prevent undue movement that could result in tip dislodgment and loss of capture, or catheter migration.Care should be taken not to kink the catheter body when securing it.In this complaint, it could not be determined if procedural factors or device handling may have contributed to the reported event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.(b)(4).
 
Event Description
It was reported that the catheter was unable to pace right after insertion.The catheter was replaced and the problem was solved.There were no patient complications reported.Further information was unable to be obtained.Patient demographic information requested but unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key6679448
MDR Text Key78789190
Report Number2015691-2017-01907
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/21/2019
Device Model NumberPE075F5
Device Lot Number60717081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-