MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS HAND SWITCH; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS HAND SWITCH; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.012VET
Device Problem Device Stops Intermittently (1599)
Patient Problem No Patient Involvement (2645)
Event Date 06/12/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and identified no failure.The device passed all testing and operational specifications.Therefore, the reported condition was not confirmed and assignable root cause was not determined.If additional information should become available, a suplemental medwatch will be submitted accordingly.
 
Event Description
This is report 2 of 3 for the same event.It was reported that during an unspecified veterinary surgical procedure, it was observed that the electric pen drive device, hand switch device and cable device stopped functioning while being used together.According to the reporter, there was a five second and then a twenty second delay and then a thirty second delay with the device.The reporter stated there were no delays to the surgical procedure.It was unknown whether a spare device was available for use.It was reported that the surgery was successfully completed.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAND SWITCH
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6682021
MDR Text Key78983611
Report Number8030965-2017-13386
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.012VET
Device Lot NumberAV21785/1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-