Model Number 304-20 |
Device Problems
Corroded (1131); High impedance (1291); Mechanical Problem (1384); Low impedance (2285)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/09/2017 |
Event Type
malfunction
|
Event Description
|
It was reported that high impedance was seen on the patient's device.The physician considered programming the patient's device off but was not sure if she should because the patient still had good seizure control.The device history records of the generator and lead were reviewed, and both devices performed to specification prior to release.The physician decided to only decrease the patient's settings since the patient wasn't having any issues associated with the high impedance and was still seizure free.There was also no known cause of the high impedance, such as trauma.No surgical intervention has occurred to date.
|
|
Event Description
|
The patient was seen again, and low impedance was present on the patient's device.The patient's device was programmed off.No surgery has occurred to date.
|
|
Event Description
|
Lead revision surgery occurred.The explanted device has not been received by the manufacturer to-date.
|
|
Manufacturer Narrative
|
Event description, corrected data: it was inadvertently not provided on follow-up report#2 that the device had been received by the manufacturer.Device available for evaluation, corrected data: the date the device was received was inadvertently not provided on follow-up report#2.
|
|
Event Description
|
The explanted device was received by the manufacturer.Analysis was completed on the returned lead portions.Visual analysis identified broken coils approximately 1.5 mm from the electrode bifurcation.Scanning electron microscopy was performed and identified the area on three of the broken coil strands as being mechanically damaged with fine pitting on one of the broken coil strands.The area on the remaining broken coil strands was identified as having evidence of a fatigue stress induced fracture with mechanical damage, no pitting and evidence of a rotational stress induced fracture which most likely completed the fracture.A flat spot was observed on the coil surface.The abraded opening found on the outer silicone tubing and the cut ends that were made during the explanted process, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.With the exception of the observed discontinuities the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The set screw marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.
|
|
Search Alerts/Recalls
|