BIOMET TRAUMA AFFIXUS HIP FRACTURE NAIL ROD, FIXATION Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467)
Patient Problem No Information (3190)
Event Date 05/30/2017
Event Type  Injury  
Manufacturer Narrative
This follow-up report is being submitted to relay additional information. Reported event was unable to be confirmed due to limited information received from the customer. Device history record (dhr) was reviewed and no discrepancies were found. Review of the complaint history determined that no further action(s) is/are required. Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
(b)(4). Customer has indicated that the product will not be returned to zimmer biomet for investigation. The investigation is in process. Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an initial hip fracture trauma procedure, while the surgeon was attempting to insert the affixus short nail following implantation of the lag screw, the targeting device missed the static hole for the distal cortical screw. The surgeon took out all of the implants and completed the surgery with a long affixus nail. A one and a half hour delay in the surgery was reported due to the malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAFFIXUS HIP FRACTURE NAIL
Type of DeviceROD, FIXATION
Manufacturer (Section D)
BIOMET TRAUMA
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6687083
MDR Text Key108782135
Report Number0001825034-2017-04544
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK100238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number814313180
Device Lot Number118900
Other Device ID NumberREFERENCE H10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A

Patient Treatment Data
Date Received: 07/05/2017 Patient Sequence Number: 1
-
-