SMITHS MEDICAL ASD, INC. CLEO® 90 INFUSION SET SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 21-7231-24
Device Problem Material Fragmentation (1261)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 06/19/2017
Event Type  malfunction  
Manufacturer Narrative
Three photos were returned for evaluation. Photographic inspection found that the cannula detachment was detected. One device was returned for evaluation in used condition without its original packaging. Visual inspection of the device found that the cannula was detached. A review of the testing and inspection documents was performed and deemed adequate and correct. A review of the manufacturing process was performed on a similar part and found no discrepancies. Three sample devices were functionally tested and detected no detachment. Based on the evidence during observation, the root cause was attributed to a manufacturing issue due to an assembly issue.
 
Manufacturer Narrative
Customer has not returned the device to the manufacturer for device evaluation. If the device becomes available and is returned and evaluated, the manufacturer will file a follow-up report detailing the results of the evaluation.
 
Event Description
It was reported that during removal of a cleo¿0 infusion set, the set "ripped apart" while still inside the patient's body. The cannula of the device "got stuck" in the patient's body, leaving about a millimeter protruding enough that the patient was able to pull it out completely with tweezers. No permanent injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCLEO® 90 INFUSION SET
Type of DeviceSET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS HEALTHCARE MFG
s.a. de c.v. ave calidad no. 4
parque industrial internaciona
tijuana, 22425
MX 22425
Manufacturer Contact
jennifer meng
6000 nathan lane north
minneapolis, MN 55442
7633833078
MDR Report Key6690847
MDR Text Key195840839
Report Number3012307300-2017-01417
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042172
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 02/07/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/20/2021
Device Catalogue Number21-7231-24
Device Lot Number76X206
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-