The customer received a questionable low ckl creatine kinase result for one patient sample.The initial result was 9 u/l and was reported outside the laboratory.The physician questioned the result because it did not agree with the clinical picture and the previous result of 1939 u/l.The sample was repeated and the result was 1565 u/l.This result was believed to be correct.The patient was not adversely affected.The reagent lot number was 20471501 with an expiration date of 11/30/2017.The customer stated the same sample generated an alarm indicating a clot when tested on another analyzer in the laboratory.The customer did not know if there was a clot in the sample at the time of the event.The field service representative determined the erroneous result was either due to a clot in the sample or the analyzer not detecting the sample was clotted.He checked the analyzer, verified the performance of the rinse mechanism, and checked all probe alignments.He ran a precision check with results well within guidelines.
|
A specific root cause could not be identified.Additional information for further investigation was requested but was not provided.Based on the provided calibration, qc, and analyzer check data, instrument or reagent problems were ruled out.Since the same sample generated an alarm indicating a clot when tested on another analyzer in the laboratory, a preanalytic issue was suspected.
|