MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® BLUE LINE ULTRA® TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® BLUE LINE ULTRA® TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/810/085
Device Problem Material Perforation (2205)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/30/2017
Event Type  malfunction  
Manufacturer Narrative
Customer has not returned the device to the manufacturer for device evaluation.If the device becomes available and is returned and evaluated, the manufacturer will file a follow-up report detailing the results of the evaluation.
 
Event Description
It was reported that the cuff of a portex® blue line ultra® tracheostomy tube "perforated".The cuff was "drilled into the patient".The fault was identified eight days after placement.It was noted that the patient was comfortable and "without respiratory effect".No injury was reported.
 
Manufacturer Narrative
One portex® blue line ultra® tracheostomy tube was received in used condition.The sample was visually inspected at a distance of 12 to 24 inches.No holes were detected.The cuff was inflated and submerged under water.A leak was found in the cuff.A review of the manufacturing and quality processes was conducted and was considered adequate and correct.Per previous complaints, retraining of the production personnel was conducted in order to reinforce the inflation test operation.Based on the evidence, the complaint was confirmed.The root cause is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® BLUE LINE ULTRA® TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS HEALTHCARE MFG
s.a. de c.v. ave calidad no. 4
parque industrial internaciona
tijuana, 22425
MX   22425
Manufacturer Contact
jennifer meng
6000 nathan lane north
minneapolis, MN 55442
7633833078
MDR Report Key6718474
MDR Text Key80215260
Report Number3012307300-2017-01505
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date03/03/2021
Device Catalogue Number100/810/085
Device Lot Number3146361
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-