U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/12/2017
Event Type  malfunction  
Event Description
The customer noted the device buttons would get hot and stick.The facility reports having another device available for use.No user injury reported.
 
Manufacturer Narrative
The device was received for evaluation.The reported complaint of the buttons getting hot and sticking could not be reproduced.Unit was tested at speed settings between 100 and 10,000 rpm¿s in forward and reverse for 5 minutes each.Also oscillate for 5 minutes in highest setting (9).Product was tested on a known good dyonics power, dii, and dii eip control units with and without footswitch.Mdu was tested using proper irrigation and 2 different blades were utilized during functional testing.All 3 buttons were exercised for 5 minutes each.At no time did mdu buttons overheat or stick.All functions perform as expected.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the functional testing process.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 w. william cannon drive
austin, TX 78735
MDR Report Key6722943
MDR Text Key80420926
Report Number1643264-2017-00416
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Device Lot NumberAAL24656
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-