• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.BRAUN SURGICAL SA DAFILON BLUE 5/0 (1) 45CM DS16; SUTURES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA DAFILON BLUE 5/0 (1) 45CM DS16; SUTURES Back to Search Results
Model Number C0932124
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Manufacturing site evaluation: evaluation on-going.
 
Event Description
Country of complaint: (b)(6).It is reported that the thread is detaching from the needle.
 
Manufacturer Narrative
Samples received: no samples received.Analysis and results: there are no previous complaints of this batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical warehouse.As no samples have been received and no units are available in b.Braun surgical, s.A.We have only reviewed the batch manufacturing record and this product had a normal process and the results during the process fulfill b.Braun surgical requirements.Needle attachment results conducted on samples during the manufacturing process were 0.58 kgf in average and 0.53 kgf in minimum and fulfilled the requirements (0.23 kgf in average and 0.11 kgf in minimum).Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyze it.Please note that when no samples are received our analysis is very limited.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 5/0 (1) 45CM DS16
Type of Device
SUTURES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
Manufacturer (Section G)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP   08191
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key6723140
MDR Text Key80415332
Report Number3003639970-2017-00373
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K1551165
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0932124
Device Catalogue NumberC0932124
Device Lot Number616376
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Distributor Facility Aware Date07/10/2017
Date Manufacturer Received06/26/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-