Catalog Number 72200616
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
Device investigation narrative - one powermax elite motor drive unit, part number 72200616 was received on july 14, 2017 and confirmed to be serial number (b)(4) as reported in the complaint. The reported complaint of shorting out could not be confirmed. Product did not short out during functional testing. Unit was tested at speed settings between 100 and 10,000 rpm¿s in forward and reverse for 5 minutes each. Also oscillate for 5 minutes in highest setting (9). Product was tested on a known good dyonics power, dii, and dii eip control units with and without footswitch. At no time did mdu short out or overheat. All functions perform as expected. The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the functional testing process. (b)(4).
 
Event Description
It was reported that the mdu shorted out during the procedure. A backup device was available to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key6726307
MDR Text Key80549580
Report Number1643264-2017-00420
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/19/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-