AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1050-040
Device Problem Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual inspection was performed on the returned device.The reported cracked sidearm was confirmed.It is likely that syringe or flushing device was over-tightened causing the sidearm to crack at the threads.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the procedure was to treat a lesion located in the right upper arm that was mildly tortuous, mildly calcifed and 70-80% stenosed.After advancement of the 5.0 x 40 mm armada 35 balloon catheter to the lesion the connection between the balloon catheter and the indeflator was observed by the naked eye to be cracked.Another armada 35 balloon catheter was used successfully.No adverse patient effects or clinically significant delay in the procedure were reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6733207
MDR Text Key80767796
Report Number2024168-2017-06098
Device Sequence Number1
Product Code LIT
UDI-Device Identifier08717648154607
UDI-Public08717648154607
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Type of Report Initial
Report Date 07/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Catalogue NumberB1050-040
Device Lot Number61219G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-