• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL MIDWEST STYLUS ATC HANDPIECE; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL MIDWEST STYLUS ATC HANDPIECE; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 882300
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Though no medical/surgical intervention was required to preclude a serious injury in this event, there have been previously reported events involving a similar device that resulted in the need for medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.Therefore, this event meets the criteria for reportability per 21 cfr part 803.
 
Event Description
In this event a doctor reported that an atc mini handpiece overheated.There was no injury or intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST STYLUS ATC HANDPIECE
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key6733587
MDR Text Key80702741
Report Number1419322-2017-00168
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K003518
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 07/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number882300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-