SYNTHES OBERDORF ELECTRIC PEN DRIVE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.010_LL |
Device Problems
Overheating of Device (1437); Device Stops Intermittently (1599); Device Inoperable (1663)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/28/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device was returned for evaluation.
Reliability engineering evaluated the device and observed that the device was not working and there were signs of an unknown substance on top of the motor.
Therefore, the reported condition was confirmed.
The assignable root cause was determined to be due to improper cleaning methods.
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
(b)(4).
The actual device has been returned and is currently pending evaluation.
Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported that during an unspecified surgical procedure, it was observed that the electric pen drive device was working intermittently.
According to the report, the device was working as expected at the beginning of the case.
The reporter stated that the device then started working intermittently and heating up.
The reporter stated that after a few more minutes, the device stopped working all together.
There was an approximate five minute delay to the surgical procedure while a spare device was retrieved.
It was reported that the spare device was used to successfully complete the surgery.
There was patient involvement.
There were no reports of injuries, medical intervention or prolonged hospitalization.
All available information has been disclosed.
If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|