• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-35
Device Problem Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/30/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on 06/27/2017 that on (b)(6) 2017 the sensor wire was missing. The attempted sensor insertion was at the abdomen. No additional event or patient information is available. No product or data were provided for evaluation. The reported missing sensor wire could not be confirmed. A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6740588
MDR Text Key80931269
Report Number3004753838-2017-42965
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation
Type of Report Initial
Report Date 06/27/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/28/2017
Device Model Number9500-35
Device Catalogue NumberSTS-GL-005
Device Lot Number5218556
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-