• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HAR36
Device Problems Thermal Decomposition of Device (1071); Smoking (1585)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/05/2017
Event Type  malfunction  
Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, the device revealed evidence of clinical use including an indentation in the black pad.The teflon pad was not attached to the jaw.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: - user expectation of adaptive tissue technology functionality.- tissue accumulation between the blade and jaw assembly.- improper usage and inadequate cleaning of instrument.The instructions for use (ifu) state: -care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this occurs, there may be an instrument failure, and the generator touchscreen displays a troubleshooting message.- for optimal performance and to avoid tissue sticking, clean the instrument blade, clamp arm, and distal end of the shaft throughout the procedure by activating the instrument tip in saline.Note: do not touch the instrument to metal while activated.Note: do not clean the blade tip with abrasives.It can be wiped with a moist gauze sponge to remove tissue, if necessary.- if tissue is still visible in the clamp arm, use hemostats to remove residue, taking care not to actuate the hand piece.If desired, the instrument may be unplugged.The reported event will continue to be monitored through post-market surveillance.
 
Event Description
It was reported that the harmonic scalpel vaporized the teflon pad inside the patient.The padding completely burned and created smoke in the abdomen.There was no patient injury or medical intervention, and extended procedure time reported was a few minutes.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer Contact
moira barton varty
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key6745854
MDR Text Key81128393
Report Number0001056128-2017-00094
Device Sequence Number1
Product Code NLQ
UDI-Device Identifier00885825017347
UDI-Public(01)00885825017347(17)180224(10)5871917
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/24/2018
Device Model NumberHAR36
Device Catalogue NumberHAR36RR
Device Lot Number5871917
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-