• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT7
Device Problem Output Problem (3005)
Patient Problem No Code Available (3191)
Event Date 07/07/2017
Event Type  Injury  
Manufacturer Narrative
Evaluation summary: the device was not returned for evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.There have been no other complaints reported in the lot number.Additional information has been requested.(b)(4).
 
Event Description
A facilities representative reported an intraocular lens (iol) was exchanged as the diopter and power of lens was too high.The originally implanted lens was a 15.0 diopter lens.It was exchanged for a same model lens 13.5 diopter.Additional information was requested.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The customer indicated the use of a qualified cartridge and handpiece.Cartridge product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.Two viscoelastics were indicated, only one is qualified for this lens with the qualified cartridge combinations.The product investigation could not identify a root cause.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6749799
MDR Text Key81241192
Report Number1119421-2017-00971
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberSN6AT7
Device Catalogue NumberSN6AT7U150
Device Lot Number12507985
Other Device ID Number00380652276351
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age42 YR
-
-