It was reported that high impedance was found on the patient's device during generator replacement surgery due to low battery.System diagnostics were not performed on the old generator, and the high impedance was seen after the existing lead was inserted into the new generator.Therefore, it could not be determined if the high impedance was present prior to the surgery.The lead was not revised at that time.Programming history from the old generator was reviewed, and did not show any evidence of high impedance prior to the generator replacement surgery.The explanted generator was discarded by the hospital.Therefore, no analysis could be performed.No further information has been received, and no lead revision has occurred to date.
|