The customer stated that they received questionable results for four patient samples tested for the elecsys afp assay (afp) and papp-a elecsys (pappa) on a cobas e 411 immunoassay analyzer (e411).Neither the papp-a reagent nor a similar reagent is available for distribution in the united states.Of the four samples, one sample had an erroneous afp result.The erroneous result was not reported outside of the laboratory.The sample initially resulted as 1.07 ug/l.The sample was repeated on (b)(6) 2017, resulting as 98.47 ug/l accompanied by a data flag.The sample was of good quality and there was no foam on the sample surface.No adverse events were alleged to have occurred with the patient.The afp reagent lot number was 17674700, with an expiration date of 10/31/2017.The field service engineer checked the customer's water quality and the analyzer liquid level detection voltage.He performed checks of the analyzer, ran performance testing, and ran blank cell calibration.Performance testing showed no issues.Precision studies were performed and were okay.
|
Quality control results prior to the event were within specifications.Upon review of the alarm trace, two liquid level detection alarms were observed on (b)(6) 2017.The pinch valves were 8 months old, which is beyond the mandatory exchange cycle.A specific root cause could not be determined based on the provided information.Since the issue occurred with multiple assays and based on the received alarm messages, a sample pipetting error most likely occurred.The pipetting error may have been caused by pre-analytic sample handling.
|