Catalog Number 532.101
Device Problem Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4). (b)(6). The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment. The device was evaluated and it was determined that the trigger of the device was blocked, jammed, and moving heavily. Therefore the reported condition was confirmed. The assignable root cause was determined to be due to wear from normal use and servicing. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the small battery drive device would stop working, and after a while it would start again. It was unknown if there was a delay in the procedure due to the event, or if a spare device was available for use. There was patient involvement reported. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of the event was not reported. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOLIBRI II
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ CH4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6756626
MDR Text Key81488391
Report Number8030965-2017-14049
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 07/18/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number532.101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2017
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-