• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616S
Device Problems Failure to Cycle (1142); Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/19/2017
Event Type  malfunction  
Event Description
During ankle surgery, the reported device was heating up and not spinning.No patient or user injury was reported.There was a brief delay reported.A back up device of the same kind was available.
 
Manufacturer Narrative
The device was received on (b)(4) 2017 for evaluation.Complaint of overheating and motor stall error was confirmed.Cause of overheating and errors is a corroded motor/gearbox.The motor/gearbox was removed from the housing.The gearbox was removed from motor.The cable assembly and motor passed functional testing.The gearbox was found to be jammed and corroded internally.The root cause of this event was identified to be associated with corrosion of the motor and gearbox assembly.A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A corrective action has been initiated to mitigate future recurrence of similar events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL,MDU, HAND CNTRL, PWRMX
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 w. william cannon drive
austin, TX 78735
MDR Report Key6760378
MDR Text Key81792404
Report Number1643264-2017-00451
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616S
Device Lot NumberAAL83048
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received07/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-